Table 3 Transducer Block Configurations (Page 14)
FILTER_FLOW_SPIKE_FILTER_PERCENTAGE Float R/W
After 'Percent Len' measurements with good quality
have been displayed a new measurement with good
quality is deemed valid and displayed when the
difference between the maximum and minimum of the
present measurement and ('Percent Len' - 1) previous
consecutive measurements is less than the
measurement range (default of 27fps for Flow) times
('Percent' / 100).
FILTER_FLOW_SPIKE_FILTER_FILT_PCT_WINDOW_LEN
Unsigned
Char
R/W
After 'Percent Len' measurements with good quality
have been displayed a new measurement with good
quality is deemed valid and displayed when the
difference between the maximum and minimum of the
present measurement and ('Percent Len' - 1) previous
consecutive measurements is less than the
measurement range (default of 27fps for Flow) times
('Percent' / 100).
GVF_NR_FILTER
FILTER_GVF_NR_FILTER_ENABLE
Unsigned
Char
R/W
Enables or disables the GVF noise reduction filter.
FILTER_GVF_NR_FILTER_MAGNITUDE_SELECTION
Unsigned
Char
R/W
Selects GVF noise reduction filter magnitude.
GVF_DAMPING_FILTER
FILTER_GVF_1ST_ORDER_DAMPING_FILTER_ENABLE
Unsigned
Char
R/W
Enables or disables GVF 1st order damping filter.
FILTER_GVF_DAMPING_TAU Float R/W
Sets tau value for GVF 1st order damping filter.
GVF_SPIKE_FILTER
FILTER_GVF_SPIKE_FILTER_ENABLE
Unsigned
Char
R/W
Enables or disables GVF spike filter.
FILTER_GVF_SPIKE_FILTER_NO_FLOW_LENGTH
Unsigned
Char
R/W
Defines the required number of consecutive
measurements with good quality before a measurement
is deemed valid and displayed.
Copyright © 2010 CiDRA Corporate Services Page 5-15
20958-01 Rev 02